HT-998XSLD Wet Leakage Test System
HT-998XSLD Wet Leakage Test Systemss is a test system developed by Hopetech for the wet insulation test of photovoltaic modules, which is applicable to the IEC61215 standard MQT15 clause and IEC61730 MST17 clause.
Overview
Criteria met as following:
IEC61215-MQT15
IEC61730-MST17
UL1703-27
This equipment is a test system developed by Hopetech for the wet insulation test of photovoltaic modules, which is applicable to the IEC61215 standard MQT15 clause and IEC61730 MST17 clause.
Temperature control system (integrated cooling and heating)
- Working principle:The refrigerant absorbs the heat of the cooled material in the evaporator and vaporizes it into steam, and the compressor continuously extracts the generated steam from the evaporator. The compressed high-temperature and high-pressure steam is sent to the condenser and then condensed into a high-pressure liquid to the cooling medium (such as water, air, etc.), which enters the evaporator after being depressurized by the throttling mechanism. It vaporizes again, absorbs the heat of the cooled object, and so on. During heating, the refrigerant changes the direction of refrigerant flow through the four-way valve, and the direction of refrigerant flow is just opposite to that of refrigeration, and the refrigerant first passes through the evaporator, then back to the condenser, and finally back to the compressor.
- Temperature range: 5-35°C
- Temperature measurement accuracy: 0.1°C;
- Compressor power: 1HP;
- Heating power: 3KW;
- Heating element: nickel-chromium wire;
- Sheath material: stainless steel;
- Temperature control accuracy: ±1°C;
- Temperature control mode: PID precise temperature control.
Sinks and Brackets
Sink size(L*W*H):2700*1600*250mm;
- Material: polypropylene (PP);
- Plate thickness:1.5mm;
- Bracket size (L*W*H):2700mm*1600mm*400mm;
Conductivity Tester
- Test range: 1.0~2000 μS/cm (equivalent to 500 Ω·cm~1000000 Ω·cm);
- Application temperature: 0~99°C;
- Reference temperature: 25°C (automatic temperature compensation if temperature deviation)
- Basic error: ± 1.0%FS;
- Temperature error: ±0.5°C(0~60°C) ;
- Repeatability:±0.2%FS;
- Stability: ±0.2%FS;
Insulation test system Hipot test
- Output voltage:0-15KV(VDC);
- Voltage accuracy: 3%;
- Voltage resolution: 1V;
- Output current: 5mA;
- Current accuracy: 3%;
- Current resolution: 0.1 μA;
- Boost time: 0.4S to 999.9S;
- Upper limit setting: 0.01-5mA;
- Lower limit setting: 0-5mA;
- Ground output:DC 10~100A, 01 μΩ~600m Ω
Insulation test
- Test voltage: 500~2500(VDC);
- Allowable error: ≤±5%;
- Boost time: 0.4S to 999.9S;
- Test resistance: 100k~99(GΩ);
- Resistance error: ± (3%+3 words) @(100kΩ-9.99GΩ)/±(20%+3 words)@(10GΩ-99GΩ);;
- Test time: 1~999(S)≤±3%;
- Discontinuity time: 5~99(S)≤±3%;
Others
- Power supply:AC380V,3KW;
- Floor area(L*W*H):1600*3200*600mm;
- Weight: 200kg;
- Water: 1 inch (DN25) tap water pipe;
- Drainage: Φ 100mm.
Features
- liquid temperature can be controlled automatically, and the temperature control system adopts the all-in-one cooling and heating machine to accurately control the water temperature
- It can monitor the conductivity and temperature of liquid;
- Equipped with insulation tester, which can test the insulation and leakage of the components;
- The insulating groove is made of PP material;
- Bottom installation universal brake casters;
- The bottom is equipped with a drain valve for easy cleaning
- Overload, over-current, over-voltage alarm.
Application
Specification
Serial number | Name | Quantity |
1 | Temperature control system | 1 |
2 | Sink & Bracket (standard PP, acrylic optional) | 1 |
3 | Conductivity tester (with conductivity standard solution) | 1 |
4 | Insulation Tester | 1 |